Serie ST50 e ST60
EMC - ELECTRICAL SAFETY TESTER
AC/DC withstand voltage test & ground bond test
Insulation resistance test (serie ST60)
Arc detection
Zero crossing turn-on, turn-off operation switch
Controllable ramp-up time
True RMS cut-off current measurement
Automatic fail indicator of alarm lamp and buzzer
VFD display
Standard PLC interface
Option RS-232,RS-485
Specifiche tecniche
AC/DC Withstanding Voltage Test
Test Time: 0.1 a 999S
Accuracy: +/- (2% + 2digits)
Insulation Resistance Test (serie ST60)
Resistance Range: 1...1000MOhm
Voltage Range: 250V, 500V, 1000V
Accuracy: +/- (5% + 5digits)
Input Voltage: 220 - 240VAC +/-10%
|

ST6050

|
Serie ST99
EMC - ELECTRICAL SAFETY TESTER
AC/DC withstanding voltage test, insulation resistance
test, ground resistance test & Leakage
Programmable test
Fast sampling rate, high efficiency
Large 320 x 240 graphic LCD
50 memory units with 8 steps in each
True RMS leakage current measurement
Standard PLC interface
Option RS-232, RS-485, USB, GPIB
Specifiche tecniche
Withstanding Voltage Test
Voltage Range: 0.050...5.000KVAC 0.050...6.000KVDC
Accuracy: ±(2%+5digits)
Test current: 0.001...40mAAC 0.001...20mADC
Accuracy: ±(2%+2digits)
Test time: 0...999.9S
Test frequency: 50/60Hz
Arc detection: 0...40mAAC 0...20mADC
Insulation Resistance Test (ST9930/ST9932)
Resistance Range: 1MΩ...9.999GΩ
Test Voltage: 100...1000VC
Accuracy: ±(2%+5V)
Test time: 0...999.9S |

ST9930
|